Test Date: 2016-02-09 11:35
Analysis date: 2016-05-26 09:57
Chip 14
X-ray Calibration - Target Zn Method Spectrum Chip C14
Center Center of Peak 59.44 ± 0.09 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 1.54
Rate Rate 232.80 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C14
Center Center of Peak 115.73 ± 0.09 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 2.45
Rate Rate 382.20 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C14
Center Center of Peak 145.83 ± 0.11 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 4.56
Rate Rate 218.90 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C14
Center Center of Peak 165.14 ± 0.16 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 6.61
Rate Rate 142.80 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C14
Slope Slope 43.518 ± 0.068 nElectrons/VCal
Offset Offset -185.442 ± 7.383 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C14
Rate Rate 232.80 kHz/cm²
NTrigs N Trig 5916440 Trigger
NHits N Hits 21485 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C14
Rate Rate 382.20 kHz/cm²
NTrigs N Trig 5916430 Trigger
NHits N Hits 35272 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C14
Rate Rate 218.90 kHz/cm²
NTrigs N Trig 5914619 Trigger
NHits N Hits 20196 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C14
Rate Rate 142.80 kHz/cm²
NTrigs N Trig 5914564 Trigger
NHits N Hits 13180 Hits
MoReWeb-v1.0.0-5-g82d9ff6 on branch psi46master