Test Date: 2016-02-09 11:35
Analysis date: 2016-05-26 09:45
Chip 14
X-ray Calibration - Target Zn Method Spectrum Chip C14
Center Center of Peak 51.86 ± 0.09 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 1.3
Rate Rate 253.20 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C14
Center Center of Peak 104.96 ± 0.08 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 2.17
Rate Rate 402.50 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C14
Center Center of Peak 132.87 ± 0.11 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 4.4
Rate Rate 231.70 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C14
Center Center of Peak 151.89 ± 0.14 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 3.45
Rate Rate 147.70 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C14
Slope Slope 46.339 ± 0.077 nElectrons/VCal
Offset Offset -4.952 ± 7.574 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C14
Rate Rate 253.20 kHz/cm²
NTrigs N Trig 5916217 Trigger
NHits N Hits 23373 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C14
Rate Rate 402.50 kHz/cm²
NTrigs N Trig 5916238 Trigger
NHits N Hits 37149 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C14
Rate Rate 231.70 kHz/cm²
NTrigs N Trig 5916299 Trigger
NHits N Hits 21389 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C14
Rate Rate 147.70 kHz/cm²
NTrigs N Trig 5914687 Trigger
NHits N Hits 13631 Hits
MoReWeb-v1.0.0-5-g82d9ff6 on branch psi46master