Test Date: 2016-02-17 14:21
Analysis date: 2016-05-26 09:08
Chip 14
X-ray Calibration - Target Zn Method Spectrum Chip C14
Center Center of Peak 52.19 ± 0.1 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 1.46
Rate Rate 233.60 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C14
Center Center of Peak 104.34 ± 0.08 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 2.24
Rate Rate 382.90 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C14
Center Center of Peak 131.82 ± 0.1 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 4.14
Rate Rate 221.50 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C14
Center Center of Peak 148.94 ± 0.14 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 6.86
Rate Rate 144.10 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C14
Slope Slope 47.162 ± 0.082 nElectrons/VCal
Offset Offset -63.002 ± 8.376 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C14
Rate Rate 233.60 kHz/cm²
NTrigs N Trig 5914993 Trigger
NHits N Hits 21558 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C14
Rate Rate 382.90 kHz/cm²
NTrigs N Trig 5915725 Trigger
NHits N Hits 35332 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C14
Rate Rate 221.50 kHz/cm²
NTrigs N Trig 5915501 Trigger
NHits N Hits 20439 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C14
Rate Rate 144.10 kHz/cm²
NTrigs N Trig 5915209 Trigger
NHits N Hits 13296 Hits