Test Date: 2016-01-22 13:53
Analysis date: 2016-05-26 08:50
Chip 14
X-ray Calibration - Target Zn Method Spectrum Chip C14
Center Center of Peak 56.7 ± 0.09 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 1.41
Rate Rate 230.30 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C14
Center Center of Peak 115.93 ± 0.09 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 2.44
Rate Rate 376.80 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C14
Center Center of Peak 147.47 ± 0.11 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 4.77
Rate Rate 219.80 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C14
Center Center of Peak 167.47 ± 0.14 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 4.4
Rate Rate 143.70 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C14
Slope Slope 41.404 ± 0.062 nElectrons/VCal
Offset Offset 52.958 ± 6.685 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C14
Rate Rate 230.30 kHz/cm²
NTrigs N Trig 5914416 Trigger
NHits N Hits 21244 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C14
Rate Rate 376.80 kHz/cm²
NTrigs N Trig 5915127 Trigger
NHits N Hits 34765 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C14
Rate Rate 219.80 kHz/cm²
NTrigs N Trig 5914793 Trigger
NHits N Hits 20280 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C14
Rate Rate 143.70 kHz/cm²
NTrigs N Trig 5915285 Trigger
NHits N Hits 13257 Hits
MoReWeb-v1.0.0-5-g82d9ff6 on branch psi46master