Test Date: 2015-12-16 10:28
Analysis date: 2016-05-26 07:18
Chip 8
X-ray Calibration - Target Zn Method Spectrum Chip C8
Center Center of Peak 45.12 ± 0.07 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 1.87
Rate Rate 294.40 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C8
Center Center of Peak 92.09 ± 0.07 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 1.82
Rate Rate 443.20 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C8
Center Center of Peak 117.19 ± 0.1 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 3.75
Rate Rate 236.90 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C8
Center Center of Peak 134.05 ± 0.12 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 3.44
Rate Rate 148.80 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C8
Slope Slope 51.646 ± 0.158 nElectrons/VCal
Offset Offset 100.074 ± 17.079 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C8
Rate Rate 294.40 kHz/cm²
NTrigs N Trig 5914630 Trigger
NHits N Hits 27160 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C8
Rate Rate 443.20 kHz/cm²
NTrigs N Trig 5914804 Trigger
NHits N Hits 40896 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C8
Rate Rate 236.90 kHz/cm²
NTrigs N Trig 5914390 Trigger
NHits N Hits 21855 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C8
Rate Rate 148.80 kHz/cm²
NTrigs N Trig 5914753 Trigger
NHits N Hits 13733 Hits
MoReWeb-v1.0.0-5-g82d9ff6 on branch psi46master