Test Date: 2015-09-10 17:49
Analysis date: 2016-05-26 05:49
Chip 14
X-ray Calibration - Target Zn Method Spectrum Chip C14
Center Center of Peak 51.98 ± 0.1 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 1.41
Rate Rate 235.70 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C14
Center Center of Peak 102.97 ± 0.09 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 2.61
Rate Rate 373.80 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C14
Center Center of Peak 130.02 ± 0.11 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 3.29
Rate Rate 217.30 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C14
Center Center of Peak 147.6 ± 0.14 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 3.65
Rate Rate 142.20 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C14
Slope Slope 48.142 ± 0.089 nElectrons/VCal
Offset Offset -102.464 ± 8.76 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C14
Rate Rate 235.70 kHz/cm²
NTrigs N Trig 5914772 Trigger
NHits N Hits 21752 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C14
Rate Rate 373.80 kHz/cm²
NTrigs N Trig 5914595 Trigger
NHits N Hits 34489 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C14
Rate Rate 217.30 kHz/cm²
NTrigs N Trig 5914645 Trigger
NHits N Hits 20054 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C14
Rate Rate 142.20 kHz/cm²
NTrigs N Trig 5914522 Trigger
NHits N Hits 13120 Hits
MoReWeb-v1.0.0-5-g82d9ff6 on branch psi46master