Test Date: 2015-09-04 17:08
Analysis date: 2016-05-26 04:27
Chip 14
X-ray Calibration - Target Zn Method Spectrum Chip C14
Center Center of Peak 49.67 ± 0.1 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 1.35
Rate Rate 212.20 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C14
Center Center of Peak 99.46 ± 0.09 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 1.76
Rate Rate 362.60 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C14
Center Center of Peak 126.23 ± 0.11 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 2.63
Rate Rate 214.00 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C14
Center Center of Peak 143.04 ± 0.14 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 2.99
Rate Rate 137.60 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C14
Slope Slope 49.119 ± 0.092 nElectrons/VCal
Offset Offset -37.022 ± 8.82 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C14
Rate Rate 212.20 kHz/cm²
NTrigs N Trig 5920170 Trigger
NHits N Hits 19600 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C14
Rate Rate 362.60 kHz/cm²
NTrigs N Trig 5920339 Trigger
NHits N Hits 33493 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C14
Rate Rate 214.00 kHz/cm²
NTrigs N Trig 5920384 Trigger
NHits N Hits 19761 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C14
Rate Rate 137.60 kHz/cm²
NTrigs N Trig 5920372 Trigger
NHits N Hits 12712 Hits
MoReWeb-v1.0.0-5-g82d9ff6 on branch psi46master