Test Date: 2015-09-04 17:08
Analysis date: 2016-05-26 04:21
Chip 8
X-ray Calibration - Target Zn Method Spectrum Chip C8
Center Center of Peak 53.27 ± 0.07 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 1.29
Rate Rate 276.80 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C8
Center Center of Peak 103.2 ± 0.07 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 2.62
Rate Rate 434.10 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C8
Center Center of Peak 130.03 ± 0.09 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 5.12
Rate Rate 233.90 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C8
Center Center of Peak 148.24 ± 0.12 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 4.42
Rate Rate 148.40 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C8
Slope Slope 48.179 ± 0.136 nElectrons/VCal
Offset Offset -115.106 ± 16.45 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C8
Rate Rate 276.80 kHz/cm²
NTrigs N Trig 5916882 Trigger
NHits N Hits 25550 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C8
Rate Rate 434.10 kHz/cm²
NTrigs N Trig 5916698 Trigger
NHits N Hits 40070 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C8
Rate Rate 233.90 kHz/cm²
NTrigs N Trig 5916436 Trigger
NHits N Hits 21586 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C8
Rate Rate 148.40 kHz/cm²
NTrigs N Trig 5917105 Trigger
NHits N Hits 13694 Hits
MoReWeb-v1.0.0-5-g82d9ff6 on branch psi46master