Test Date: 2015-09-04 17:08
Analysis date: 2016-05-26 03:09
Chip 8
X-ray Calibration - Target Zn Method Spectrum Chip C8
Center Center of Peak 56.24 ± 0.07 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 1.59
Rate Rate 355.90 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C8
Center Center of Peak 109.48 ± 0.07 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 2.94
Rate Rate 498.70 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C8
Center Center of Peak 137.51 ± 0.09 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 4.34
Rate Rate 253.90 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C8
Center Center of Peak 154.87 ± 0.13 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 4.01
Rate Rate 160.60 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C8
Slope Slope 46.198 ± 0.061 nElectrons/VCal
Offset Offset -199.496 ± 6.167 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C8
Rate Rate 355.90 kHz/cm²
NTrigs N Trig 5920860 Trigger
NHits N Hits 32872 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C8
Rate Rate 498.70 kHz/cm²
NTrigs N Trig 5920183 Trigger
NHits N Hits 46060 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C8
Rate Rate 253.90 kHz/cm²
NTrigs N Trig 5920083 Trigger
NHits N Hits 23448 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C8
Rate Rate 160.60 kHz/cm²
NTrigs N Trig 5921085 Trigger
NHits N Hits 14839 Hits
MoReWeb-v1.0.0-5-g82d9ff6 on branch psi46master