Test Date: 2015-09-04 17:08
Analysis date: 2016-05-26 03:09
Chip 14
X-ray Calibration - Target Zn Method Spectrum Chip C14
Center Center of Peak 51.69 ± 0.09 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 1.7
Rate Rate 241.80 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C14
Center Center of Peak 104.18 ± 0.08 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 2.96
Rate Rate 389.30 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C14
Center Center of Peak 131.61 ± 0.1 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 4.61
Rate Rate 225.20 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C14
Center Center of Peak 149.12 ± 0.13 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 4.58
Rate Rate 145.50 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C14
Slope Slope 46.959 ± 0.076 nElectrons/VCal
Offset Offset -30.398 ± 7.592 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C14
Rate Rate 241.80 kHz/cm²
NTrigs N Trig 5920860 Trigger
NHits N Hits 22334 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C14
Rate Rate 389.30 kHz/cm²
NTrigs N Trig 5920183 Trigger
NHits N Hits 35952 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C14
Rate Rate 225.20 kHz/cm²
NTrigs N Trig 5920083 Trigger
NHits N Hits 20801 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C14
Rate Rate 145.50 kHz/cm²
NTrigs N Trig 5921085 Trigger
NHits N Hits 13443 Hits
MoReWeb-v1.0.0-5-g82d9ff6 on branch psi46master