Test Date: 2015-08-24 17:49
Analysis date: 2016-05-26 02:42
Chip 14
X-ray Calibration - Target Zn Method Spectrum Chip C14
Center Center of Peak 50.95 ± 0.09 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 1.57
Rate Rate 207.90 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C14
Center Center of Peak 103.37 ± 0.07 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 2.8
Rate Rate 363.40 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C14
Center Center of Peak 131.1 ± 0.1 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 4.89
Rate Rate 212.10 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C14
Center Center of Peak 148.89 ± 0.12 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 4.62
Rate Rate 138.20 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C14
Slope Slope 46.865 ± 0.076 nElectrons/VCal
Offset Offset 11.594 ± 7.512 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C14
Rate Rate 207.90 kHz/cm²
NTrigs N Trig 5919287 Trigger
NHits N Hits 19197 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C14
Rate Rate 363.40 kHz/cm²
NTrigs N Trig 5919329 Trigger
NHits N Hits 33557 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C14
Rate Rate 212.10 kHz/cm²
NTrigs N Trig 5919702 Trigger
NHits N Hits 19591 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C14
Rate Rate 138.20 kHz/cm²
NTrigs N Trig 5920199 Trigger
NHits N Hits 12759 Hits
MoReWeb-v1.0.0-5-g82d9ff6 on branch psi46master