Test Date: 2015-08-13 17:13
Analysis date: 2016-05-26 01:23
Chip 9
X-ray Calibration - Target Zn Method Spectrum Chip C9
Center Center of Peak 54.63 ± 0.07 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 3.45
Rate Rate 345.60 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C9
Center Center of Peak 110.9 ± 0.07 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 6.45
Rate Rate 483.50 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C9
Center Center of Peak 141.16 ± 0.1 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 9.55
Rate Rate 257.90 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C9
Center Center of Peak 159.75 ± 0.13 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 9.08
Rate Rate 170.60 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C9
Slope Slope 43.488 ± 0.056 nElectrons/VCal
Offset Offset 26.181 ± 5.645 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C9
Rate Rate 345.60 kHz/cm²
NTrigs N Trig 5919890 Trigger
NHits N Hits 31913 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C9
Rate Rate 483.50 kHz/cm²
NTrigs N Trig 5920122 Trigger
NHits N Hits 44652 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C9
Rate Rate 257.90 kHz/cm²
NTrigs N Trig 5920418 Trigger
NHits N Hits 23815 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C9
Rate Rate 170.60 kHz/cm²
NTrigs N Trig 5920404 Trigger
NHits N Hits 15752 Hits
MoReWeb-v1.0.0-5-g82d9ff6 on branch psi46master