Test Date: 2015-08-11 15:16
Analysis date: 2016-05-26 01:05
Chip 9
X-ray Calibration - Target Zn Method Spectrum Chip C9
Center Center of Peak 51.09 ± 0.09 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 1.28
Rate Rate 263.50 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C9
Center Center of Peak 105.83 ± 0.07 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 2.69
Rate Rate 425.60 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C9
Center Center of Peak 133.59 ± 0.11 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 4.0
Rate Rate 229.90 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C9
Center Center of Peak 151.87 ± 0.14 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 3.9
Rate Rate 145.70 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C9
Slope Slope 46.969 ± 0.143 nElectrons/VCal
Offset Offset -115.717 ± 17.278 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C9
Rate Rate 263.50 kHz/cm²
NTrigs N Trig 5919835 Trigger
NHits N Hits 24338 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C9
Rate Rate 425.60 kHz/cm²
NTrigs N Trig 5919714 Trigger
NHits N Hits 39300 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C9
Rate Rate 229.90 kHz/cm²
NTrigs N Trig 5919593 Trigger
NHits N Hits 21234 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C9
Rate Rate 145.70 kHz/cm²
NTrigs N Trig 5919350 Trigger
NHits N Hits 13456 Hits
MoReWeb-v1.0.0-5-g82d9ff6 on branch psi46master