Test Date: 2015-08-11 15:16
Analysis date: 2016-05-26 01:05
Chip 8
X-ray Calibration - Target Zn Method Spectrum Chip C8
Center Center of Peak 47.83 ± 0.1 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 1.58
Rate Rate 281.30 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C8
Center Center of Peak 99.89 ± 0.08 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 2.41
Rate Rate 437.90 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C8
Center Center of Peak 127.35 ± 0.11 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 3.54
Rate Rate 238.10 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C8
Center Center of Peak 145.2 ± 0.13 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 3.9
Rate Rate 151.00 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C8
Slope Slope 47.226 ± 0.085 nElectrons/VCal
Offset Offset 139.893 ± 8.176 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C8
Rate Rate 281.30 kHz/cm²
NTrigs N Trig 5919835 Trigger
NHits N Hits 25977 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C8
Rate Rate 437.90 kHz/cm²
NTrigs N Trig 5919714 Trigger
NHits N Hits 40440 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C8
Rate Rate 238.10 kHz/cm²
NTrigs N Trig 5919593 Trigger
NHits N Hits 21985 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C8
Rate Rate 151.00 kHz/cm²
NTrigs N Trig 5919350 Trigger
NHits N Hits 13942 Hits