Test Date: 2015-08-11 09:36
Analysis date: 2016-05-26 00:23
Chip 14
X-ray Calibration - Target Zn Method Spectrum Chip C14
Center Center of Peak 57.21 ± 0.09 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 1.08
Rate Rate 212.60 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C14
Center Center of Peak 112.69 ± 0.09 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 2.72
Rate Rate 360.90 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C14
Center Center of Peak 141.83 ± 0.12 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 3.85
Rate Rate 213.30 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C14
Center Center of Peak 161.81 ± 0.16 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 7.41
Rate Rate 139.90 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C14
Slope Slope 44.357 ± 0.073 nElectrons/VCal
Offset Offset -139.314 ± 7.572 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C14
Rate Rate 212.60 kHz/cm²
NTrigs N Trig 5919763 Trigger
NHits N Hits 19629 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C14
Rate Rate 360.90 kHz/cm²
NTrigs N Trig 5919868 Trigger
NHits N Hits 33325 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C14
Rate Rate 213.30 kHz/cm²
NTrigs N Trig 5920407 Trigger
NHits N Hits 19696 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C14
Rate Rate 139.90 kHz/cm²
NTrigs N Trig 5919841 Trigger
NHits N Hits 12916 Hits
MoReWeb-v1.0.0-5-g82d9ff6 on branch psi46master