Test Date: 2015-08-11 09:36
Analysis date: 2016-05-26 00:11
Chip 9
X-ray Calibration - Target Zn Method Spectrum Chip C9
Center Center of Peak 45.24 ± 0.07 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 1.94
Rate Rate 286.80 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C9
Center Center of Peak 91.41 ± 0.06 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 2.15
Rate Rate 429.10 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C9
Center Center of Peak 115.69 ± 0.09 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 4.38
Rate Rate 232.40 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C9
Center Center of Peak 132.27 ± 0.11 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 4.85
Rate Rate 147.60 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C9
Slope Slope 53.282 ± 0.08 nElectrons/VCal
Offset Offset -12.17 ± 6.845 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C9
Rate Rate 286.80 kHz/cm²
NTrigs N Trig 5916485 Trigger
NHits N Hits 26468 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C9
Rate Rate 429.10 kHz/cm²
NTrigs N Trig 5916982 Trigger
NHits N Hits 39612 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C9
Rate Rate 232.40 kHz/cm²
NTrigs N Trig 5917223 Trigger
NHits N Hits 21450 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C9
Rate Rate 147.60 kHz/cm²
NTrigs N Trig 5917492 Trigger
NHits N Hits 13625 Hits
MoReWeb-v1.0.0-5-g82d9ff6 on branch psi46master