Test Date: 2015-08-18 09:38
Analysis date: 2016-05-25 23:59
Chip 13
X-ray Calibration - Target Zn Method Spectrum Chip C13
Center Center of Peak 51.78 ± 0.12 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 1.46
Rate Rate 174.40 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C13
Center Center of Peak 102.73 ± 0.1 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 1.88
Rate Rate 311.60 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C13
Center Center of Peak 129.38 ± 0.13 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 2.64
Rate Rate 181.70 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C13
Center Center of Peak 147.22 ± 0.15 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 2.84
Rate Rate 115.60 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C13
Slope Slope 48.686 ± 0.187 nElectrons/VCal
Offset Offset -145.465 ± 22.694 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C13
Rate Rate 174.40 kHz/cm²
NTrigs N Trig 5919899 Trigger
NHits N Hits 16104 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C13
Rate Rate 311.60 kHz/cm²
NTrigs N Trig 5920075 Trigger
NHits N Hits 28776 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C13
Rate Rate 181.70 kHz/cm²
NTrigs N Trig 5920083 Trigger
NHits N Hits 16782 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C13
Rate Rate 115.60 kHz/cm²
NTrigs N Trig 5920444 Trigger
NHits N Hits 10681 Hits
MoReWeb-v1.0.0-5-g82d9ff6 on branch psi46master