Test Date: 2015-08-12 14:10
Analysis date: 2016-05-25 23:35
Chip 14
X-ray Calibration - Target Zn Method Spectrum Chip C14
Center Center of Peak 50.6 ± 0.09 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 1.73
Rate Rate 254.20 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C14
Center Center of Peak 106.09 ± 0.08 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 2.66
Rate Rate 398.40 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C14
Center Center of Peak 134.69 ± 0.11 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 5.24
Rate Rate 226.80 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C14
Center Center of Peak 153.29 ± 0.14 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 4.75
Rate Rate 146.60 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C14
Slope Slope 45.755 ± 0.143 nElectrons/VCal
Offset Offset -0.119 ± 17.696 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C14
Rate Rate 254.20 kHz/cm²
NTrigs N Trig 5919137 Trigger
NHits N Hits 23476 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C14
Rate Rate 398.40 kHz/cm²
NTrigs N Trig 5919335 Trigger
NHits N Hits 36785 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C14
Rate Rate 226.80 kHz/cm²
NTrigs N Trig 5918984 Trigger
NHits N Hits 20938 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C14
Rate Rate 146.60 kHz/cm²
NTrigs N Trig 5919057 Trigger
NHits N Hits 13534 Hits
MoReWeb-v1.0.0-5-g82d9ff6 on branch psi46master