Test Date: 2015-07-10 14:30
Analysis date: 2016-05-25 21:43
Chip 14
X-ray Calibration - Target Zn Method Spectrum Chip C14
Center Center of Peak 54.34 ± 0.08 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 27.37
Rate Rate 263.20 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C14
Center Center of Peak 104.49 ± 0.07 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 14.28
Rate Rate 407.70 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C14
Center Center of Peak 132.23 ± 0.11 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 8.95
Rate Rate 230.20 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C14
Center Center of Peak 150.79 ± 0.12 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 7.54
Rate Rate 151.50 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C14
Slope Slope 46.782 ± 0.131 nElectrons/VCal
Offset Offset -32.394 ± 15.926 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C14
Rate Rate 263.20 kHz/cm²
NTrigs N Trig 5918636 Trigger
NHits N Hits 24300 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C14
Rate Rate 407.70 kHz/cm²
NTrigs N Trig 5918796 Trigger
NHits N Hits 37640 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C14
Rate Rate 230.20 kHz/cm²
NTrigs N Trig 5918668 Trigger
NHits N Hits 21256 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C14
Rate Rate 151.50 kHz/cm²
NTrigs N Trig 5919004 Trigger
NHits N Hits 13991 Hits
MoReWeb-v1.0.0-5-g82d9ff6 on branch psi46master