Test Date: 2015-07-10 14:30
Analysis date: 2016-05-25 21:41
Chip 14
X-ray Calibration - Target Zn Method Spectrum Chip C14
Center Center of Peak 50.57 ± 0.09 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 19.97
Rate Rate 244.30 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C14
Center Center of Peak 102.69 ± 0.07 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 11.77
Rate Rate 393.00 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C14
Center Center of Peak 130.83 ± 0.11 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 8.32
Rate Rate 227.90 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C14
Center Center of Peak 148.73 ± 0.13 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 6.96
Rate Rate 149.00 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C14
Slope Slope 46.968 ± 0.065 nElectrons/VCal
Offset Offset 26.794 ± 6.83 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C14
Rate Rate 244.30 kHz/cm²
NTrigs N Trig 5918723 Trigger
NHits N Hits 22561 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C14
Rate Rate 393.00 kHz/cm²
NTrigs N Trig 5919185 Trigger
NHits N Hits 36285 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C14
Rate Rate 227.90 kHz/cm²
NTrigs N Trig 5919263 Trigger
NHits N Hits 21045 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C14
Rate Rate 149.00 kHz/cm²
NTrigs N Trig 5918883 Trigger
NHits N Hits 13760 Hits