Test Date: 2015-07-07 08:06
Analysis date: 2016-05-25 21:35
Chip 14
X-ray Calibration - Target Zn Method Spectrum Chip C14
Center Center of Peak 45.65 ± 0.15 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 37.47
Rate Rate 196.20 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C14
Center Center of Peak 97.13 ± 0.06 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 22.56
Rate Rate 355.30 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C14
Center Center of Peak 124.23 ± 0.1 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 12.63
Rate Rate 207.40 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C14
Center Center of Peak 142.15 ± 0.13 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 9.01
Rate Rate 136.60 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C14
Slope Slope 48.055 ± 0.135 nElectrons/VCal
Offset Offset 187.618 ± 14.917 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C14
Rate Rate 196.20 kHz/cm²
NTrigs N Trig 5919755 Trigger
NHits N Hits 18120 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C14
Rate Rate 355.30 kHz/cm²
NTrigs N Trig 5919616 Trigger
NHits N Hits 32815 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C14
Rate Rate 207.40 kHz/cm²
NTrigs N Trig 5915229 Trigger
NHits N Hits 19141 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C14
Rate Rate 136.60 kHz/cm²
NTrigs N Trig 5919108 Trigger
NHits N Hits 12610 Hits
MoReWeb-v1.0.0-5-g82d9ff6 on branch psi46master