Test Date: 2015-08-21 17:53
Analysis date: 2016-05-25 20:58
Chip 14
X-ray Calibration - Target Zn Method Spectrum Chip C14
Center Center of Peak 48.55 ± 0.07 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 1.54
Rate Rate 249.10 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C14
Center Center of Peak 96.78 ± 0.07 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 2.57
Rate Rate 388.10 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C14
Center Center of Peak 121.91 ± 0.09 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 4.71
Rate Rate 225.20 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C14
Center Center of Peak 138.22 ± 0.13 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 4.78
Rate Rate 143.50 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C14
Slope Slope 52.078 ± 0.165 nElectrons/VCal
Offset Offset -186.266 ± 18.539 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C14
Rate Rate 249.10 kHz/cm²
NTrigs N Trig 5920255 Trigger
NHits N Hits 23007 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C14
Rate Rate 388.10 kHz/cm²
NTrigs N Trig 5920062 Trigger
NHits N Hits 35843 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C14
Rate Rate 225.20 kHz/cm²
NTrigs N Trig 5920270 Trigger
NHits N Hits 20796 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C14
Rate Rate 143.50 kHz/cm²
NTrigs N Trig 5921025 Trigger
NHits N Hits 13253 Hits
MoReWeb-v1.0.0-5-g82d9ff6 on branch psi46master