Test Date: 2016-11-01 11:04
Analysis date: 2016-11-01 11:25
Chip 15
X-ray Calibration - Target Zn Method Spectrum Chip C15
Center Center of Peak 71.21 ± 4.71 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 1.75
Rate Rate 54.20 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C15
Center Center of Peak 127.97 ± 1.39 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 2.24
Rate Rate 126.70 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C15
Center Center of Peak 151.77 ± 1.07 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 2.09
Rate Rate 70.80 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C15
Center Center of Peak 172.95 ± 1.59 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 1.32
Rate Rate 42.80 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C15
Slope Slope 40.767 ± 3.68 nElectrons/VCal
Offset Offset -30.867 ± 580.719 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C15
Rate Rate 54.20 kHz/cm²
NTrigs N Trig 2989207 Trigger
NHits N Hits 2529 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C15
Rate Rate 126.70 kHz/cm²
NTrigs N Trig 2990547 Trigger
NHits N Hits 5909 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C15
Rate Rate 70.80 kHz/cm²
NTrigs N Trig 2990272 Trigger
NHits N Hits 3301 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C15
Rate Rate 42.80 kHz/cm²
NTrigs N Trig 2990332 Trigger
NHits N Hits 1996 Hits
MoReWeb-v1.0.5-18-g6921b8e on branch L1Xray