Test Date: 2016-11-02 17:42
Analysis date: 2016-11-02 17:59
Chip 14
X-ray Calibration - Target Zn Method Spectrum Chip C14
Center Center of Peak 70.99 ± 0.62 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 1.1
Rate Rate 42.20 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C14
Center Center of Peak 117.05 ± 0.68 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 1.98
Rate Rate 114.50 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C14
Center Center of Peak 140.71 ± 0.79 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 1.24
Rate Rate 65.60 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C14
Center Center of Peak 153.58 ± 1.3 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 1.49
Rate Rate 40.80 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C14
Slope Slope 57.547 ± 1.95 nElectrons/VCal
Offset Offset -1894.293 ± 256.206 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C14
Rate Rate 42.20 kHz/cm²
NTrigs N Trig 2990069 Trigger
NHits N Hits 1970 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C14
Rate Rate 114.50 kHz/cm²
NTrigs N Trig 2989800 Trigger
NHits N Hits 5339 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C14
Rate Rate 65.60 kHz/cm²
NTrigs N Trig 2989927 Trigger
NHits N Hits 3062 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C14
Rate Rate 40.80 kHz/cm²
NTrigs N Trig 2990058 Trigger
NHits N Hits 1901 Hits