Test Date: 2016-08-18 13:47
Analysis date: 2016-08-18 15:25
Chip 14
X-ray Calibration - Target Zn Method Spectrum Chip C14
Center Center of Peak 54.78 ± 0.18 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 1.15
Rate Rate 119.10 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C14
Center Center of Peak 107.75 ± 0.11 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 1.53
Rate Rate 362.30 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C14
Center Center of Peak 135.46 ± 0.13 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 4.57
Rate Rate 219.70 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C14
Center Center of Peak 154.23 ± 0.17 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 4.74
Rate Rate 147.90 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C14
Slope Slope 46.526 ± 0.106 nElectrons/VCal
Offset Offset -152.832 ± 12.616 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C14
Rate Rate 119.10 kHz/cm²
NTrigs N Trig 5922735 Trigger
NHits N Hits 11003 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C14
Rate Rate 362.30 kHz/cm²
NTrigs N Trig 5921746 Trigger
NHits N Hits 33472 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C14
Rate Rate 219.70 kHz/cm²
NTrigs N Trig 5921818 Trigger
NHits N Hits 20294 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C14
Rate Rate 147.90 kHz/cm²
NTrigs N Trig 5921735 Trigger
NHits N Hits 13659 Hits
MoReWeb-v1.0.2-10-gbc52a17 on branch psi46master