Test Date: 2015-12-11 15:01
Analysis date: 2015-12-15 09:52
Chip 1
X-ray Calibration - Target Zn Method Spectrum Chip C1
Center Center of Peak 52.95 ± 0.17 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 0.94
Rate Rate 102.10 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C1
Center Center of Peak 106.3 ± 0.1 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 2.03
Rate Rate 334.90 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C1
Center Center of Peak 134.59 ± 0.12 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 3.72
Rate Rate 203.20 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C1
Center Center of Peak 153.11 ± 0.15 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 3.34
Rate Rate 136.00 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C1
Slope Slope 46.015 ± 0.117 nElectrons/VCal
Offset Offset -36.383 ± 12.94 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C1
Rate Rate 102.10 kHz/cm²
NTrigs N Trig 5918532 Trigger
NHits N Hits 9428 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C1
Rate Rate 334.90 kHz/cm²
NTrigs N Trig 5918593 Trigger
NHits N Hits 30919 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C1
Rate Rate 203.20 kHz/cm²
NTrigs N Trig 5919288 Trigger
NHits N Hits 18762 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C1
Rate Rate 136.00 kHz/cm²
NTrigs N Trig 5919049 Trigger
NHits N Hits 12555 Hits
MoReWeb-v0.6.7-111-gcc5e703 on branch 20151208_Readback