Test Date: 2015-12-11 15:01
Analysis date: 2015-12-15 09:48
Chip 9
X-ray Calibration - Target Zn Method Spectrum Chip C9
Center Center of Peak 52.08 ± 0.16 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 1.22
Rate Rate 135.70 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C9
Center Center of Peak 104.72 ± 0.09 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 1.65
Rate Rate 399.50 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C9
Center Center of Peak 131.35 ± 0.11 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 3.39
Rate Rate 223.40 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C9
Center Center of Peak 148.32 ± 0.15 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 3.34
Rate Rate 141.90 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C9
Slope Slope 49.414 ± 0.181 nElectrons/VCal
Offset Offset -322.36 ± 22.139 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C9
Rate Rate 135.70 kHz/cm²
NTrigs N Trig 5924608 Trigger
NHits N Hits 12542 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C9
Rate Rate 399.50 kHz/cm²
NTrigs N Trig 5923908 Trigger
NHits N Hits 36919 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C9
Rate Rate 223.40 kHz/cm²
NTrigs N Trig 5924048 Trigger
NHits N Hits 20643 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C9
Rate Rate 141.90 kHz/cm²
NTrigs N Trig 5923855 Trigger
NHits N Hits 13116 Hits
MoReWeb-v0.6.7-111-gcc5e703 on branch 20151208_Readback