Test Date: 2015-12-11 15:01
Analysis date: 2015-12-15 09:48
Chip 13
X-ray Calibration - Target Zn Method Spectrum Chip C13
Center Center of Peak 48.75 ± 0.18 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 1.1
Rate Rate 72.10 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C13
Center Center of Peak 96.58 ± 0.1 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 1.4
Rate Rate 271.10 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C13
Center Center of Peak 121.88 ± 0.13 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 2.72
Rate Rate 165.00 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C13
Center Center of Peak 138.02 ± 0.15 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 3.0
Rate Rate 110.80 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C13
Slope Slope 51.37 ± 0.156 nElectrons/VCal
Offset Offset -105.293 ± 15.533 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C13
Rate Rate 72.10 kHz/cm²
NTrigs N Trig 5924608 Trigger
NHits N Hits 6665 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C13
Rate Rate 271.10 kHz/cm²
NTrigs N Trig 5923908 Trigger
NHits N Hits 25051 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C13
Rate Rate 165.00 kHz/cm²
NTrigs N Trig 5924048 Trigger
NHits N Hits 15247 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C13
Rate Rate 110.80 kHz/cm²
NTrigs N Trig 5923855 Trigger
NHits N Hits 10236 Hits
MoReWeb-v0.6.7-111-gcc5e703 on branch 20151208_Readback