Test Date: 2015-12-11 15:01
Analysis date: 2015-12-15 09:48
Chip 12
X-ray Calibration - Target Zn Method Spectrum Chip C12
Center Center of Peak 53.27 ± 0.1 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 1.23
Rate Rate 123.30 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C12
Center Center of Peak 104.78 ± 0.08 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 2.01
Rate Rate 401.00 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C12
Center Center of Peak 132.26 ± 0.11 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 3.92
Rate Rate 226.00 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C12
Center Center of Peak 150.22 ± 0.15 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 3.63
Rate Rate 145.40 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C12
Slope Slope 47.611 ± 0.074 nElectrons/VCal
Offset Offset -135.629 ± 7.936 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C12
Rate Rate 123.30 kHz/cm²
NTrigs N Trig 5924608 Trigger
NHits N Hits 11395 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C12
Rate Rate 401.00 kHz/cm²
NTrigs N Trig 5923908 Trigger
NHits N Hits 37056 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C12
Rate Rate 226.00 kHz/cm²
NTrigs N Trig 5924048 Trigger
NHits N Hits 20885 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C12
Rate Rate 145.40 kHz/cm²
NTrigs N Trig 5923855 Trigger
NHits N Hits 13437 Hits
MoReWeb-v0.6.7-111-gcc5e703 on branch 20151208_Readback