Test Date: 2015-12-11 15:01
Analysis date: 2015-12-15 09:48
Chip 1
X-ray Calibration - Target Zn Method Spectrum Chip C1
Center Center of Peak 51.85 ± 0.16 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 0.95
Rate Rate 96.10 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C1
Center Center of Peak 103.37 ± 0.09 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 1.99
Rate Rate 332.00 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C1
Center Center of Peak 129.73 ± 0.12 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 3.45
Rate Rate 205.90 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C1
Center Center of Peak 147.96 ± 0.15 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 5.98
Rate Rate 134.00 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C1
Slope Slope 48.75 ± 0.177 nElectrons/VCal
Offset Offset -180.955 ± 21.422 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C1
Rate Rate 96.10 kHz/cm²
NTrigs N Trig 5924608 Trigger
NHits N Hits 8880 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C1
Rate Rate 332.00 kHz/cm²
NTrigs N Trig 5923908 Trigger
NHits N Hits 30685 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C1
Rate Rate 205.90 kHz/cm²
NTrigs N Trig 5924048 Trigger
NHits N Hits 19024 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C1
Rate Rate 134.00 kHz/cm²
NTrigs N Trig 5923855 Trigger
NHits N Hits 12384 Hits
MoReWeb-v0.6.7-111-gcc5e703 on branch 20151208_Readback