Test Date: 2015-12-11 15:01
Analysis date: 2016-01-07 14:17
Chip 8
X-ray Calibration - Target Zn Method Spectrum Chip C8
Center Center of Peak 54.99 ± 0.11 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 1.33
Rate Rate 173.00 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C8
Center Center of Peak 106.29 ± 0.07 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 2.2
Rate Rate 452.90 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C8
Center Center of Peak 133.64 ± 0.11 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 3.11
Rate Rate 241.20 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C8
Center Center of Peak 151.32 ± 0.14 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 3.24
Rate Rate 153.10 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C8
Slope Slope 47.786 ± 0.092 nElectrons/VCal
Offset Offset -226.124 ± 9.63 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C8
Rate Rate 173.00 kHz/cm²
NTrigs N Trig 5924881 Trigger
NHits N Hits 15990 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C8
Rate Rate 452.90 kHz/cm²
NTrigs N Trig 5924563 Trigger
NHits N Hits 41857 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C8
Rate Rate 241.20 kHz/cm²
NTrigs N Trig 5924973 Trigger
NHits N Hits 22294 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C8
Rate Rate 153.10 kHz/cm²
NTrigs N Trig 5924385 Trigger
NHits N Hits 14148 Hits
MoReWeb-v0.6.7-126-g20f70e8 on branch dev-v0.7.0