Test Date: 2015-12-11 15:01
Analysis date: 2016-01-07 14:17
Chip 12
X-ray Calibration - Target Zn Method Spectrum Chip C12
Center Center of Peak 47.93 ± 0.1 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 0.93
Rate Rate 141.60 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C12
Center Center of Peak 94.77 ± 0.07 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 2.2
Rate Rate 430.00 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C12
Center Center of Peak 120.2 ± 0.09 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 3.82
Rate Rate 234.60 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C12
Center Center of Peak 136.85 ± 0.13 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 3.53
Rate Rate 147.60 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C12
Slope Slope 51.345 ± 0.16 nElectrons/VCal
Offset Offset -11.497 ± 17.65 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C12
Rate Rate 141.60 kHz/cm²
NTrigs N Trig 5924881 Trigger
NHits N Hits 13086 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C12
Rate Rate 430.00 kHz/cm²
NTrigs N Trig 5924563 Trigger
NHits N Hits 39744 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C12
Rate Rate 234.60 kHz/cm²
NTrigs N Trig 5924973 Trigger
NHits N Hits 21681 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C12
Rate Rate 147.60 kHz/cm²
NTrigs N Trig 5924385 Trigger
NHits N Hits 13640 Hits
MoReWeb-v0.6.7-126-g20f70e8 on branch dev-v0.7.0