Test Date: 2015-11-23 11:06
Analysis date: 2015-11-23 18:02
Chip 8
X-ray Calibration - Target Zn Method Spectrum Chip C8
Center |
Center of Peak |
54.69 ± 0.13 |
Vcal |
TargetEnergy |
Energy of target Zn |
8638.91 |
eV |
TargetNElectrons |
Energy of target Zn |
2399.7 |
nElectrons |
Chi2PerNDF |
Chi^2 per NDF |
1.16 |
|
Rate |
Rate |
151.50 |
kHz/cm² |
X-ray Calibration - Target Mo Method Spectrum Chip C8
Center |
Center of Peak |
112.15 ± 0.09 |
Vcal |
TargetEnergy |
Energy of target Mo |
17479.37 |
eV |
TargetNElectrons |
Energy of target Mo |
4855.38 |
nElectrons |
Chi2PerNDF |
Chi^2 per NDF |
1.81 |
|
Rate |
Rate |
407.30 |
kHz/cm² |
X-ray Calibration - Target Ag Method Spectrum Chip C8
Center |
Center of Peak |
143.02 ± 0.12 |
Vcal |
TargetEnergy |
Energy of target Ag |
22162.92 |
eV |
TargetNElectrons |
Energy of target Ag |
6156.37 |
nElectrons |
Chi2PerNDF |
Chi^2 per NDF |
5.3 |
|
Rate |
Rate |
224.30 |
kHz/cm² |
X-ray Calibration - Target Sn Method Spectrum Chip C8
Center |
Center of Peak |
162.69 ± 0.17 |
Vcal |
TargetEnergy |
Energy of target Sn |
25271.36 |
eV |
TargetNElectrons |
Energy of target Sn |
7019.82 |
nElectrons |
Chi2PerNDF |
Chi^2 per NDF |
5.49 |
|
Rate |
Rate |
144.10 |
kHz/cm² |
X-ray Calibration - Vcal Calibration Method Spectrum Chip C8
Slope |
Slope |
42.654 ± 0.072 |
nElectrons/VCal |
Offset |
Offset |
67.989 ± 8.592 |
nElectrons |
X-ray Hit Map - Target Zn Method Spectrum Chip C8
Rate |
Rate |
151.50 |
kHz/cm² |
NTrigs |
N Trig |
5925603 |
Trigger |
NHits |
N Hits |
14004 |
Hits |
X-ray Hit Map - Target Mo Method Spectrum Chip C8
Rate |
Rate |
407.30 |
kHz/cm² |
NTrigs |
N Trig |
5924900 |
Trigger |
NHits |
N Hits |
37648 |
Hits |
X-ray Hit Map - Target Ag Method Spectrum Chip C8
Rate |
Rate |
224.30 |
kHz/cm² |
NTrigs |
N Trig |
5924611 |
Trigger |
NHits |
N Hits |
20727 |
Hits |
X-ray Hit Map - Target Sn Method Spectrum Chip C8
Rate |
Rate |
144.10 |
kHz/cm² |
NTrigs |
N Trig |
5924081 |
Trigger |
NHits |
N Hits |
13321 |
Hits |
MoReWeb-v0.6.7-73-g24fed82