Test Date: 2015-11-23 11:06
Analysis date: 2015-11-23 17:26
Chip 1
X-ray Calibration - Target Zn Method Spectrum Chip C1
Center Center of Peak 53.05 ± 0.12 Vcal
TargetEnergy Energy of target Zn 8638.91 eV
TargetNElectrons Energy of target Zn 2399.7 nElectrons
Chi2PerNDF Chi^2 per NDF 1.08
Rate Rate 146.30 kHz/cm²
X-ray Calibration - Target Mo Method Spectrum Chip C1
Center Center of Peak 106.46 ± 0.09 Vcal
TargetEnergy Energy of target Mo 17479.37 eV
TargetNElectrons Energy of target Mo 4855.38 nElectrons
Chi2PerNDF Chi^2 per NDF 2.02
Rate Rate 364.50 kHz/cm²
X-ray Calibration - Target Ag Method Spectrum Chip C1
Center Center of Peak 135.01 ± 0.11 Vcal
TargetEnergy Energy of target Ag 22162.92 eV
TargetNElectrons Energy of target Ag 6156.37 nElectrons
Chi2PerNDF Chi^2 per NDF 4.08
Rate Rate 218.50 kHz/cm²
X-ray Calibration - Target Sn Method Spectrum Chip C1
Center Center of Peak 153.38 ± 0.15 Vcal
TargetEnergy Energy of target Sn 25271.36 eV
TargetNElectrons Energy of target Sn 7019.82 nElectrons
Chi2PerNDF Chi^2 per NDF 3.42
Rate Rate 144.30 kHz/cm²
X-ray Calibration - Vcal Calibration Method Spectrum Chip C1
Slope Slope 45.854 ± 0.09 nElectrons/VCal
Offset Offset -30.297 ± 9.558 nElectrons
X-ray Hit Map - Target Zn Method Spectrum Chip C1
Rate Rate 146.30 kHz/cm²
NTrigs N Trig 5926290 Trigger
NHits N Hits 13524 Hits
X-ray Hit Map - Target Mo Method Spectrum Chip C1
Rate Rate 364.50 kHz/cm²
NTrigs N Trig 5925768 Trigger
NHits N Hits 33694 Hits
X-ray Hit Map - Target Ag Method Spectrum Chip C1
Rate Rate 218.50 kHz/cm²
NTrigs N Trig 5925736 Trigger
NHits N Hits 20201 Hits
X-ray Hit Map - Target Sn Method Spectrum Chip C1
Rate Rate 144.30 kHz/cm²
NTrigs N Trig 5926432 Trigger
NHits N Hits 13342 Hits
MoReWeb-v0.6.7-73-g24fed82 on branch dev-v0.7.0